Quality Control (Field Congruence and Beam Perpendicularity Tests) For Diagnostic X-Ray Machines in Misurata City

Authors

DOI:

https://doi.org/10.54361/ajmas.261142

Keywords:

Quality Control, Field Congruence, Beam Perpendicularity, Diagnostic X-Ray, Radiation Protection

Abstract

Diagnostic radiology is a cornerstone of modern clinical medicine; however, it constitutes a primary artificial source of ionising radiation exposure to the general population. Maintaining high diagnostic image quality while adhering to radiation protection standards relies fundamentally on periodic quality control (QC) testing of radiographic systems. This study evaluates the operational performance of conventional diagnostic X-ray machines across selected public healthcare facilities in Misurata city by assessing light-to-X-ray field congruence (collimation alignment) and central ray beam perpendicularity. Six diagnostic X-ray units distributed across four healthcare centres were tested using standardized commercial beam and collimator alignment tools at a source-to-image receptor distance (SID) of 100 cm. Compliance was benchmarked against international regulatory tolerances: ≤ 2% of the SID for single-axis field congruence, < 4% for combined orthogonal axes, and < 3° for central ray angular deviation. All six evaluated units achieved a 100% pass rate across both tests. Single-axis displacements ranged from -1.50 cm to +0.50 cm (within the 2.0 cm limit), while combined orthogonal deviations ranged from 1.08 cm to 3.35 cm (within the 4.0 cm limit). Furthermore, central ray angular deviations remained within acceptable thresholds (0.5° to < 3°). These findings confirm robust mechanical calibration and high geometric accuracy across the surveyed units, directly supporting diagnostic reliability and upholding the ALARA (As Low as Reasonably Achievable) principle by minimizing patient radiation doses from unnecessary retakes and improper collimation.

Author Biography

Mohamed Almajdub, Department of Diagnostic and Therapeutic Radiology, Faculty of Medical Technology, Misurata, Libya

Dean of the Faculty of Medical Technology, Misurata, Libya. 

Lecturer at the Department of diagnostic and therapeutic radiology.

Downloads

Published

2026-11-19

How to Cite

1.
Ghabag A, Almajdub M, Abufalgha K, Abeed H, Khalleefah Y. Quality Control (Field Congruence and Beam Perpendicularity Tests) For Diagnostic X-Ray Machines in Misurata City. Alq J Med App Sci [Internet]. 2026 Nov. 19 [cited 2026 Nov. 19];:3847-52. Available from: https://journal.utripoli.edu.ly/index.php/Alqalam/article/view/2139